- Home - Site Map - Links - Contact Us - About Us - Get In The Loop Search our site:  
  394
Circular Chart Recorder with Annotation  


Input Board
General    
Input types   dc Volts, dc millivolts,
    dc milliamps (with shunt), Thermocouple, 2/3-wire RTD
      Contact closure (not chan. 1) >500ms
Input type mix     Freely configurable
Maximum number of inputs   6
Input ranges     -38 to + 38 mV;
      - 150 to +150 mV;
      -1V to + 1 V;
      -10 to + 10 V
Termination     Terminal block
Noise rejection (48 to 62Hz)    
  Common mode:   >130dB (channel to channel and channel to ground)
  Series mode:   >60dB
Maximum common mode voltage   250V continuous
Maximum series mode voltage   45mV at lowest range; 12Vpeak at highest range
Isolation (dc to 65 Hz; BS EN61010)   Installation cat. II; Pollution degree 2
Channel - to - channel:   300V RMS or dc (double insulation)
Channel - to - common electronics:   300V RMS or dc (double insulation)
Channel - to - ground:   300V RMS or dc (basic insulation)
Dielectric strength (BS EN61010)   (1 minute type tests)
Channel to channel   2300 Vac
Channel to ground   1350 Vac
Insulation resistance     >10 MW at 500V dc
Input resistance     38mV, 150 mV, 1 V ranges: >10MΩ; 10V range: 68.8 kΩ
Over voltage protection (max)   42V RMS (between I and V-);
50V RMS (I and V+) or (V+ and V-)
Open circuit detection drive   ± 57nA max
Recognition time     500ms
Minimum break resistance   10MΩ
 
DC Input ranges
Shunt/Attenuator     Internally mounted resistor modules
Additional error due to shunt   0.1% of input
Additional error due to attenuator   0.2% of input
Leakage current (Max) 38mV range:   1.7nA
Other ranges:     8nA
Performance     See table 1
 
High Resolution   Error at 20°C Temperature coefficient (per °C)
±38mV 1.4µV Typical 0.035% input + 0.030% range 37ppm of input + 1.03 ppm range
Max 0.085% input + 0.051% range 80ppm of input + 18.6 ppm range
±150mV 5.5µV Typical 0.035% input + 0.027% range 35ppm of input + 0.52 ppm range
Max 0.084% input + 0.038% range 80ppm of input + 7.8 ppm range
±1V 37µV Typical 0.035% input + 0.024% range 35ppm of input + 0.16 ppm range
Max 0.084% input + 0.029% range 80ppm of input + 1.6 ppm range
±10V 370µV Typical 0.076% input + 0.024% range 76ppm of input + 0.35 ppm range
Max 0.275% input + 0.030% range 272ppm of input + 3.5 ppm range
Table 1 DC Performance
 
Thermocouple data
Temperature scale     ITS 90
Bias current (maximum)     1.7nA
Cold junction types     Off, internal, external, remote
CJ error     1°C max; instrument at 20°C
CJ rejection ratio     50:1 minimum
Remote CJ     Via any user-defined channel
Upscale/downscale drive   High, Low or None selectable for each thermocouple channel. Channels can be any mix of High and None or Low and None, but High and Low cannot be mixed.
Types and ranges     See table 2
 
T/C Type Overall range (°C) Standard Max linearization error
B 0 To + 1820 IEC 584.1 0 to 400°C; 1.7° 400 to 1820°C; 0.03°C
C
D
E
G2
J
K
L
0 to 2300
0 to 2495
-270 to + 1000
-0 to + 2315
-210 to + 1200
-270 to + 1372
-200 to + 900
Hoskins
Hoskins
IEC 584.1
Hoskins
IEC 584.1
IEC 584.1
DIN43700:1985 (TO IPTS68)
0.12°C
0.08°C
0.03°C
0.07°C
0.02°C
0.04°C
0.20
N
R
S
T
U
NI/NiMo
Plantinel
270 to + 1300
-50 to + 1768
-50 to + 1768
-270 to + 400
-200 to + 600
0 to 1406
0 to + 1370
IEC 584.1
IEC 584.1
IEC 584.1
IEC 584.1
DIN43710:1985
Ipsen
Engelhard
0.04°C
0.04°C
0.04°C
0.02°C
0.08°C
0.14°C
0.02°C
Table 2 Thermocouple types and ranges
 
Resistance inputs
Temperature scale     ITS90
Ranges (including lead resistance)   0 to 150Ω, 0 to 600Ω, 0 to 6kΩ
Influence of lead resistance    
  Error:   Negligible (3-wire connection)
  Mismatch:   1Ω/Ω
Wetting current     250mA typical
Resolution and accuracy     See table 3
RTD types and ranges     See table 4
 
Range Ω Resolution   Error at 20°C Temperature coefficient (per °C)
0 to 150 5mΩ Typical 0.030% input + 0.047% range 20ppm of input + 2.04 ppm range
Max 0.045% input + 0.141% range 35ppm of input + 36.6 ppm range
0 to 600 22mΩ Typical 0.030% input + 0.036% range 20ppm of input + 0.97 ppm range
Max 0.045% input + 0.069% range 35ppm of input + 14.6 ppm range
0 to 6k 148mΩ Typical 0.034% input + 0.026% range 20ppm of input + 0.19 ppm range
Max 0.049% input + 0.032% range 35ppm of input + 1.9 ppm range
Table 3 Resistance ranges resolution and accuracy
 
RTD Type Overall range (°C) Standard Max linearization error
Cu10 -20 to + 400 General Electric company 0.02°C
JPT100 -220 to + 630 JIS C1604;1989 0.01°C
NI100 -60 to + 250 DIN43760;1987 0.01°C
NI120 -50 to + 170 DIN43760;1987 0.01°C
Pt100 -200 to + 850 IEC 751 0.01°C
Pt100A -200 to + 600 Eurotherm Recorders SA 0.09°C
Pt1000 -200 to + 850 IEC 751 0.01°C
Table 4 RTD types, ranges and accuracies
 
Board types and hardware options
  Standard:   Universal input/control board
  Options:   Changeover relay output board
      Analog output board
      Communications board
      Transmitter Power Supply
      PID controller
      Case heater
      PC Card
 
Environmental Performance
Temperature limits Operation:   0 to 50°C (options can reduce maximum temp.)
  Storage:   –20 to + 70°C
Humidity limits (non-condensing)   10 to 90%
Protection Standard:   NEMA3 (IP54)
  Waterproof:   NEMA4 (IP65)
Shock     BS EN60873 and BS EN61010
Vibration (BS EN60873)     1g peak at 60Hz to 150Hz
Altitude (max.)     <2000m
 
Electromagnetic compatibility (EMC)
  Emissions:   BS EN50081-2
  Immunity:   BS EN50082-2
 
Electrical safety (BS EN61010)
      Installation cat. II; Pollution degree 2
 
Physical
Bezel size     14.18" (360mm) H x 14.96" (380mm) (When viewed from the front, offset .2" (5mm) right with respect to cutout centerline
Panel cutout dimensions     13.4" (340.5mm) H x 13.58" (345mm) W (both – 0 + 1mm)
Depth behind bezel rear face   5.9" (150mm)
Weight     3.18lb (7kg) typical
Panel mounting     +5 to -30 degrees from vertical (+ = top overhangs)
 
Printing system
Pen type     Four color cartridge
Trace resolution lateral:   0.2mm
  time:   0.36mm at outer edge of chart
Default trace colors     See table 5
Printhead life Channel:   1.5 x 10 6 dots black
      1.0 x 10 6 dots (other colors)
Update rate     2Hz (1Hz for complex configurations)
Print rate (max)     1 pass every 5s
Text characters per line     39
 
Paper transport
Type     Stepper motor
Chart type     Circular
Chart speeds Menu selectable:   12, 24, 48, 72 hours or 7 days/rev
  User enterable:   1 to 960 hours/rev
 
Power requirements
Line voltage Standard:   90 to 264V at 45 to 65Hz
Low voltage option:   20 to 53 V ac/dc (ac frequency range: 45 to 400Hz)
Power (Max)     < 100VA (20VA typical)
Fuse type     Not user serviceable
Interrupt protection Standard:   40ms at 75% max instrument load
  Enhanced:   120ms at 75% max instrument load
 
Serial Communications
Type     RS 485 MODBUS ® RTU
Isolation†     Terminals to ground (100V RMS/dc (basic insulation))
 
Math pack
Number of derived channels   16
Level 1 functions     Off, constant, add, subtract, multiply, divide, modulus
Level 2 functions (additional to level 1)   See table 6
 
Channel average DV group continuous min Sample and hold Zirconia probe
Channel minimum Dv group latching max Square root High select
Channel maximum DV group continuous max 3rd order polynomial Low select
Rolling average e' Fvalue Switch
Rate of change log Relative humidity Stopwatch
DV group average 10' Linear mass flow Time stamp
DV grp latching min log Square root mass flow O correction Percentile
Table 6 Level two math functions
 
Customer linearization tables
Nº of tables available     One
Nº of point pairs     32
 
Relay outputs
Maximum switching power*   500VA or 60W
Maximum breaking current*   2 Amps within above power ratings
Maximum contact voltage*   250V within above power ratings
Isolation†      
Contact to contact:   300V RMS or dc (double insulation)
Contact to ground:   300V RMS or dc (basic insulation)
Estimated life with resistive loads*   30,000,000 operations
* With inductive loads, derate according to the graph, in which: contact life = resistive life x F1 or F2 where F1 = measured on representative examples and F2 = typical values according to experience.  
 
Analog (retransmission) outputs
Output ranges (user configurable)    
  Voltage:   0 to 10V (Source 6.3 mA max)
  Current:   0 to 20mA (max voltage drop =18V)
Update rate     1Hz
Step response (10% to 90%)   250ms
Linearity (maximum error)   0.02% of hardware range
Performance     See table 7
Isolation†      
Channel to channel:   300V RMS or dc (double insulation)
Channel to ground:   300V RMS or dc (basic insulation)
 
Range Ω   Error at 20°C Temperature coefficient (per °C)
0 to 10V Typical 5.7mV + 0.08 of output 100µV/°C +50ppm/°C of output
Max 11.mV 0 0.18 of output 300µV/°C + 70ppm/°C of output
0 to 20mA Typical 15.µA + 0.11 out output 0.2µA/°C + 50ppm/°C of output
Max 30.5µA + 0.21 of output 1µA/°C + 80ppm/°C of output
 
Transmitter Power Supply
Output voltage     3 or 6 x 25Vdc (nom) outputs
Isolation†      
Channel to channel:   100V RMS or dc (double insulation)
Channel to ground:   100V RMS or dc (basic isolation)
 
Controllers
Number     Up to 2
Type     Eurotherm type 2216 PID temperature controllers
† All Isolation figures are: DC to 65Hz; BS EN61010 Installation category II; Polution degree 2
Last Updated Tuesday, July 11, 2006