| Input Board |
| General |
|
|
| Input types |
|
dc Volts, dc millivolts, |
| |
|
dc milliamps (with shunt), Thermocouple, 2/3-wire RTD |
| |
|
|
Contact closure (not chan. 1) >500ms |
| Input type mix |
|
|
Freely configurable |
| Maximum number of inputs |
|
6 |
| Input ranges |
|
|
-38 to + 38 mV; |
| |
|
|
- 150 to +150 mV; |
| |
|
|
-1V to + 1 V; |
| |
|
|
-10 to + 10 V |
| Termination |
|
|
Terminal block |
| Noise rejection (48 to 62Hz) |
|
|
| |
Common mode: |
|
>130dB (channel to channel and channel to ground) |
| |
Series mode: |
|
>60dB |
| Maximum common mode voltage |
|
250V continuous |
| Maximum series mode voltage |
|
45mV at lowest range; 12Vpeak at highest range |
| Isolation (dc to 65 Hz; BS EN61010) |
|
Installation cat. II; Pollution degree 2 |
| Channel - to - channel: |
|
300V RMS or dc (double insulation) |
| Channel - to - common electronics: |
|
300V RMS or dc (double insulation) |
| Channel - to - ground: |
|
300V RMS or dc (basic insulation) |
| Dielectric strength |
(BS EN61010) |
|
(1 minute type tests) |
| Channel to channel |
|
2300 Vac |
| Channel to ground |
|
1350 Vac |
| Insulation resistance |
|
|
>10 MW at 500V dc |
| Input resistance |
|
|
38mV, 150 mV, 1 V ranges: >10MΩ; 10V range: 68.8 kΩ |
| Over voltage protection (max) |
|
42V RMS (between I and V-); 50V RMS (I and V+) or (V+ and V-) |
| Open circuit detection drive |
|
± 57nA max |
| Recognition time |
|
|
500ms |
| Minimum break resistance |
|
10MΩ |
| |
| DC Input ranges |
| Shunt/Attenuator |
|
|
Internally mounted resistor modules |
| Additional error due to shunt |
|
0.1% of input |
| Additional error due to attenuator |
|
0.2% of input |
| Leakage current (Max) 38mV range: |
|
1.7nA |
| Other ranges: |
|
|
8nA |
| Performance |
|
|
See table 1 |
| |
| High |
Resolution |
|
Error at 20°C |
Temperature coefficient (per °C) |
| ±38mV |
1.4µV |
Typical |
0.035% input +
0.030% range |
37ppm of input +
1.03 ppm range |
| Max |
0.085% input +
0.051% range |
80ppm of input +
18.6 ppm range |
| ±150mV |
5.5µV |
Typical |
0.035% input +
0.027% range |
35ppm of input +
0.52 ppm range |
| Max |
0.084% input +
0.038% range |
80ppm of input +
7.8 ppm range |
| ±1V |
37µV |
Typical |
0.035% input +
0.024% range |
35ppm of input +
0.16 ppm range |
| Max |
0.084% input +
0.029% range |
80ppm of input +
1.6 ppm range |
| ±10V |
370µV |
Typical |
0.076% input +
0.024% range |
76ppm of input +
0.35 ppm range |
| Max |
0.275% input +
0.030% range |
272ppm of input +
3.5 ppm range |
|
|
Table 1 DC Performance |
| |
| Thermocouple data |
| Temperature scale |
|
|
ITS 90 |
| Bias current (maximum) |
|
|
1.7nA |
| Cold junction types |
|
|
Off, internal, external, remote |
| CJ error |
|
|
1°C max; instrument at 20°C |
| CJ rejection ratio |
|
|
50:1 minimum |
| Remote CJ |
|
|
Via any user-defined channel |
| Upscale/downscale drive |
|
High, Low or None selectable for each thermocouple channel. Channels can be any mix of High and None or Low and None, but High and Low cannot be mixed. |
| Types and ranges |
|
|
See table 2 |
| |
| T/C Type |
Overall range
(°C) |
Standard |
Max
linearization error |
| B |
0 To + 1820 |
IEC 584.1 |
0 to 400°C; 1.7°
400 to 1820°C; 0.03°C |
C D E G2 J K L |
0 to 2300 0 to 2495 -270 to + 1000 -0 to + 2315 -210 to + 1200 -270 to + 1372 -200 to + 900 |
Hoskins Hoskins IEC 584.1 Hoskins IEC 584.1 IEC 584.1 DIN43700:1985 (TO IPTS68) |
0.12°C 0.08°C 0.03°C 0.07°C 0.02°C 0.04°C 0.20 |
N R S T U NI/NiMo Plantinel |
270 to + 1300 -50 to + 1768 -50 to + 1768 -270 to + 400 -200 to + 600 0 to 1406 0 to + 1370 |
IEC 584.1 IEC 584.1 IEC 584.1 IEC 584.1 DIN43710:1985 Ipsen Engelhard |
0.04°C 0.04°C 0.04°C 0.02°C 0.08°C 0.14°C 0.02°C |
|
|
Table 2 Thermocouple types and ranges |
| |
| Resistance inputs |
| Temperature scale |
|
|
ITS90 |
| Ranges (including lead resistance) |
|
0 to 150Ω, 0 to 600Ω, 0 to 6kΩ |
| Influence of lead resistance |
|
|
| |
Error: |
|
Negligible (3-wire connection) |
| |
Mismatch: |
|
1Ω/Ω |
| Wetting current |
|
|
250mA typical |
| Resolution and accuracy |
|
|
See table 3 |
| RTD types and ranges |
|
|
See table 4 |
| |
| Range Ω |
Resolution |
|
Error at 20°C |
Temperature coefficient (per °C) |
| 0 to
150 |
5mΩ |
Typical |
0.030% input +
0.047% range |
20ppm of input +
2.04 ppm range |
| Max |
0.045% input +
0.141% range |
35ppm of input +
36.6 ppm range |
| 0 to
600 |
22mΩ |
Typical |
0.030% input +
0.036% range |
20ppm of input +
0.97 ppm range |
| Max |
0.045% input +
0.069% range |
35ppm of input +
14.6 ppm range |
| 0 to 6k |
148mΩ |
Typical |
0.034% input +
0.026% range |
20ppm of input +
0.19 ppm range |
| Max |
0.049% input +
0.032% range |
35ppm of input +
1.9 ppm range |
|
|
Table 3 Resistance ranges resolution and accuracy |
| |
| RTD Type |
Overall range
(°C) |
Standard |
Max
linearization error |
| Cu10 |
-20 to + 400 |
General Electric company |
0.02°C |
| JPT100 |
-220 to + 630 |
JIS C1604;1989 |
0.01°C |
| NI100 |
-60 to + 250 |
DIN43760;1987 |
0.01°C |
| NI120 |
-50 to + 170 |
DIN43760;1987 |
0.01°C |
| Pt100 |
-200 to + 850 |
IEC 751 |
0.01°C |
| Pt100A |
-200 to + 600 |
Eurotherm Recorders SA |
0.09°C |
| Pt1000 |
-200 to + 850 |
IEC 751 |
0.01°C |
|
|
Table 4 RTD types, ranges and accuracies |
| |
| Board types and hardware options |
| |
Standard: |
|
Universal input/control board |
| |
Options: |
|
Changeover relay output board |
| |
|
|
Analog output board |
| |
|
|
Communications board |
| |
|
|
Transmitter Power Supply |
| |
|
|
PID controller |
| |
|
|
Case heater |
| |
|
|
PC Card |
| |
| Environmental Performance |
| Temperature limits |
Operation: |
|
0 to 50°C (options can reduce maximum temp.) |
| |
Storage: |
|
–20 to + 70°C |
| Humidity limits (non-condensing) |
|
10 to 90% |
| Protection |
Standard: |
|
NEMA3 (IP54) |
| |
Waterproof: |
|
NEMA4 (IP65) |
| Shock |
|
|
BS EN60873 and BS EN61010 |
| Vibration (BS EN60873) |
|
|
1g peak at 60Hz to 150Hz |
| Altitude (max.) |
|
|
<2000m |
| |
| Electromagnetic compatibility (EMC) |
| |
Emissions: |
|
BS EN50081-2 |
| |
Immunity: |
|
BS EN50082-2 |
| |
| Electrical safety (BS EN61010) |
| |
|
|
Installation cat. II; Pollution degree 2 |
| |
| Physical |
| Bezel size |
|
|
14.18" (360mm) H x 14.96" (380mm) (When viewed from the front, offset .2" (5mm) right with respect to cutout centerline |
| Panel cutout dimensions |
|
|
13.4" (340.5mm) H x 13.58" (345mm) W (both – 0 + 1mm) |
| Depth behind bezel rear face |
|
5.9" (150mm) |
| Weight |
|
|
3.18lb (7kg) typical |
| Panel mounting |
|
|
+5 to -30 degrees from vertical (+ = top overhangs) |
| |
| Printing system |
| Pen type |
|
|
Four color cartridge |
| Trace resolution |
lateral: |
|
0.2mm |
| |
time: |
|
0.36mm at outer edge of chart |
| Default trace colors |
|
|
See table 5 |
| Printhead life |
Channel: |
|
1.5 x 10 6 dots black |
| |
|
|
1.0 x 10 6 dots (other colors) |
| Update rate |
|
|
2Hz (1Hz for complex configurations) |
| Print rate (max) |
|
|
1 pass every 5s |
| Text characters per line |
|
|
39 |
| |
| Paper transport |
| Type |
|
|
Stepper motor |
| Chart type |
|
|
Circular |
| Chart speeds |
Menu selectable: |
|
12, 24, 48, 72 hours or 7 days/rev |
| |
User enterable: |
|
1 to 960 hours/rev |
| |
| Power requirements |
| Line voltage |
Standard: |
|
90 to 264V at 45 to 65Hz |
| Low voltage option: |
|
20 to 53 V ac/dc (ac frequency range: 45 to 400Hz) |
| Power (Max) |
|
|
< 100VA (20VA typical) |
| Fuse type |
|
|
Not user serviceable |
| Interrupt protection |
Standard: |
|
40ms at 75% max instrument load |
| |
Enhanced: |
|
120ms at 75% max instrument load |
| |
| Serial Communications |
| Type |
|
|
RS 485 MODBUS ® RTU |
| Isolation† |
|
|
Terminals to ground (100V RMS/dc (basic insulation)) |
| |
| Math pack |
| Number of derived channels |
|
16 |
| Level 1 functions |
|
|
Off, constant, add, subtract, multiply, divide, modulus |
| Level 2 functions (additional to level 1) |
|
See table 6 |
| |
| Channel average |
DV group continuous min |
Sample and hold |
Zirconia probe |
| Channel minimum |
Dv group latching max |
Square root |
High select |
| Channel maximum |
DV group continuous max |
3rd order polynomial |
Low select |
| Rolling average |
e' |
Fvalue |
Switch |
| Rate of change |
log |
Relative humidity |
Stopwatch |
| DV group average |
10' |
Linear mass flow |
Time stamp |
| DV grp latching min |
log |
Square root mass flow |
O correction Percentile |
|
|
Table 6 Level two math functions |
| |
| Customer linearization tables |
| Nº of tables available |
|
|
One |
| Nº of point pairs |
|
|
32 |
| |
| Relay outputs |
| Maximum switching power* |
|
500VA or 60W |
| Maximum breaking current* |
|
2 Amps within above power ratings |
| Maximum contact voltage* |
|
250V within above power ratings |
| Isolation† |
|
|
|
| Contact to contact: |
|
300V RMS or dc (double insulation) |
| Contact to ground: |
|
300V RMS or dc (basic insulation) |
| Estimated life with resistive loads* |
|
30,000,000 operations |
| * With inductive loads, derate according
to the graph, in which:
contact life = resistive life x F1 or F2
where F1 = measured on representative
examples and F2 = typical values
according to experience. |
|
 |
| |
| Analog (retransmission) outputs |
| Output ranges (user configurable) |
|
|
| |
Voltage: |
|
0 to 10V (Source 6.3 mA max) |
| |
Current: |
|
0 to 20mA (max voltage drop =18V) |
| Update rate |
|
|
1Hz |
| Step response (10% to 90%) |
|
250ms |
| Linearity (maximum error) |
|
0.02% of hardware range |
| Performance |
|
|
See table 7 |
| Isolation† |
|
|
|
| Channel to channel: |
|
300V RMS or dc (double insulation) |
| Channel to ground: |
|
300V RMS or dc (basic insulation) |
| |
| Range Ω |
|
Error at 20°C |
Temperature coefficient (per °C) |
| 0 to
10V |
Typical |
5.7mV + 0.08 of
output |
100µV/°C +50ppm/°C
of output |
| Max |
11.mV 0 0.18 of
output |
300µV/°C + 70ppm/°C
of output |
| 0 to
20mA |
Typical |
15.µA + 0.11 out
output |
0.2µA/°C + 50ppm/°C
of output |
| Max |
30.5µA + 0.21 of
output |
1µA/°C + 80ppm/°C
of output |
|
| |
| Transmitter Power Supply |
| Output voltage |
|
|
3 or 6 x 25Vdc (nom) outputs |
| Isolation† |
|
|
|
| Channel to channel: |
|
100V RMS or dc (double insulation) |
| Channel to ground: |
|
100V RMS or dc (basic isolation) |
| |
| Controllers |
| Number |
|
|
Up to 2 |
| Type |
|
|
Eurotherm type 2216 PID temperature controllers |
|
† All Isolation figures are: DC to 65Hz; BS EN61010 Installation category II; Polution degree 2 |