- Home - Site Map - Links - Contact Us - About Us - Get In The Loop Search our site:  
5000B
Technical Specifications
 
Environmental performance    
Temperature limits Operation:   0 to 50°C
  Storage:   –20 to + 70°C
Humidity limits Operation/Storage:   5% to 95% RH (non condensing)
Protection     IP20
Shock     BS EN61010
Vibration (BS EN60068-2-6 Test Fc)   2g peak
Altitude     < 2000 meters

Clock (RTC) data    
Temperature stability    
  0 to 40°C   –3 to +2 ppm
  –40 to +85°C   ±7.5 ppm
Ageing     ±1 ppm per year

Electromagnetic compatibility (EMC)
Emissions and immunity   BS EN61326

Electrical Safety    
(BS EN61010)     Installation category II; Pollution degree 2

Physical      
Mounting method     DIN rail (T35) or wall mounted.
Mounting angle     Connectors to be horizontal for T/C inputs - otherwise no constraints.
Weight     < 1.5kg

Operator interface    
Full operation, configuration and file transfer from remote PC

Power requirements    
Supply voltage     18 to 30V dc
Maximum power drain   10 Watts
Inrush current (maximum)   18 Amps
      Eurotherm Model 5000P can be used

Back-up Battery    
Type     Poly-carbonmonofluoride/lithium (BR2330) Part No. PA261095
Support time (RTC)     1 year min. with recorder unpowered
Replacement period     3 years
Stored data     Time; date; values for totalisers, counters and timers; batch data; Fvalue, Rolling average, Stopwatch etc.

Update/archive rates    
Input/relay output sample rate   8Hz (all channels)
Display update     Network dependent
Sample value     Value at sample time
Trend value     Latest value at display update time
Calculations     8Hz update of all alarm setpoints, maths, totaliser, counter etc. values

Inputs      
General      
Input types     dc Volts, dc millivolts,
      dc milliamps (needs external shunt)
      Thermocouple, 2/3 wire RTD, Contact closure (not ch1 or 7)
      >60msec
Input type mix     Freely configurable
Maximum number of inputs   12
A/D conversion method   >16 bits, 2nd order delta-sigma
Termination     Edge connector / Terminal block
Noise rejection Common mode:   >140dB (channel-to-channel)
      >140dB (channel-to-ground)
  Series mode:   >60dB
Maximum common mode voltage   250 Volts continuous
Maximum series mode voltage   45mV at lowest (38mV) range
      12 Volts at highest (10V) range
Isolation* Channel-to-channel:   300V RMS or dc (double insulation)
Channel-to-common electronics:   300V RMS or dc (double insulation)
  Channel-to-ground:   300V RMS or dc (basic insulation)
Dielectric strength (BS EN61010)   (1 minute type tests)
  Channel-to-channel:   2500 Volts ac
  Channel-to-ground:   1500 Volts ac
Insulation resistance   >10MΩ at 500 Volts dc
Input impedance 10 Volt range:   68.8kΩ
  All other ranges:   >10MΩ
Overvoltage protection   50V peak (150V peak with attenuator)
Open circuit     Recognition time: 500 msec
Current:     57nA
Minimum break resistance:   10MΩ

DC input ranges    
Shunt type     Externally mounted resistor modules
Additional error due to shunt   0.1% of input
Additional error due to attenuator   0.2% of input
 
Low Range High Range Resolution Maximum error
(Instrument at 20°C)
Worst case temperature
performance
–38mV 38mV 1.4μV 0.085% input + 0.051% range 80ppm of input per °C
–150mV 150mV 5.5μV 0.084% input + 0.038% range 80ppm of input per °C
–1V 1V 37μV 0.084% input + 0.029% range 80ppm of input per °C
–10V 10V 370μV 0.275% input + 0.030% range 272ppm of input per °C
DC ranges and performance

Thermocouple data    
Temperature scale     ITS90
Bias current 0.05nA
Cold junction Types: Off, internal, external, remote
Error: 1°C max. with instrument at 25°C
Rejection ratio: 50:1 minimum
Upscale/downscale drive High, low or none selectable for each thermocouple channel
  Additional error   0.01°C (typ.) if High/low selected
 
T/C Type Overall range (°C) Standard Max linearization error
B 0to +1820 IEC 584.1 0 to 400°C=1.7°C
400 to 1820°C=0.03°C
C 0to +2300 Hoskins 0.12°C
D 0to +2495 Hoskins 0.08°C
E –270 to +1000 IEC 584.1 0.03°C
G2 0 to +2315 Hoskins 0.07°C
J –210 to +1200 IEC 584.1 0.02°C
K –270 to +1372 IEC 584.1 0.04°C
L –200 to +900 DIN43700:1985 (To IPTS68) 0.20°C
N –270 to +1300 IEC 584.1 0.04°C
R –50 to +1768 IEC 584.1 0.04°C
S –50 to +1768 IEC 584.1 0.04°C
T –270 to +400 IEC 584.1 0.02°C
U –200 to +600 DIN43700:1985 0.04°C
NiMoNiCo –50 to +1410 ASTM E1751-95 0.06°C
NiNiMo 0 to +1406 Ipsen 0.14°C
Platinel 0 to +1370 Engelhard 0.02°C
Pt20%Rh /Pt40%Rh 0 to +1888 ASTM E1751-95 0.07°C
Thermocouple types and ranges

Resistance inputs    
Temperature scale     ITS90
Influence of lead resistance    
  Error:   Negligible
  Mismatch:   1Ω/Ω
       
Low Range High Range Resolution Maximum error
(Instrument at 20°C)
Worst case temperature
performance
150Ω 5mΩ 0.045% input + 0.110% range 35ppm of input per °C
600Ω 22mΩ 0.045% input + 0.065% range 35ppm of input per °C
6000Ω 148mΩ 0.049% input + 0.035% range 35ppm of input per °C
Resistance ranges – accuracy and resolution
       
RTD Type Overall range (°C) Standard Max linearization error (4102C, 4102M only)
Cu10 –20 to +400 General Electric Co. 0.02°C
Cu53 –70 to +200 RC21-4-1966 <0.01°C
JPT100 –220 to +630 JIS C 1604:1989 0 01°C
Ni1000 –60 to +250 DIN43760:1987 0.01°C
Ni120 –50 to +170 DIN43760:1987 0.01°C
Pt100 –200 to +850 IEC 751 0.01°C
Pt100A –200 to +600 Eurotherm Recorders SA 0.09°C
Pt1000 –200 to +850 IEC 751 0.01°C
Resistance thermometer types and ranges

Relay Outputs      
Number of relays fitted    
  Standard:   1
  Optional:   Up to 2 boards, each with 3 relays
Termination     Edge connector / Terminal block
Maximum ac switching power*   500VA
Maximum ac breaking current*   2A within above power ratings
Maximum ac contact voltage*   250V within above power ratings
Isolation† Relay-to-relay:   300V RMS or dc (double insulation)
  Relay-to-ground:   300V RMS or dc (basic insulation)
* With resistive loads. With inductive loads, derate according to Graph 1, in which:
Contact life = resistive life x F1 or F2 where
F1 = measured on representative examples and
F2 = typical values according to experience
       

Event Input      
Number of inputs     6 discrete inputs
Maximum No. of boards   2
Isolation Event input to ground:   100V RMS or dc (double insulation)
Event input to Event input:   0V
Recognition levels Low:   –30V to +0.8V
  High:   2 to 30V
Maximum frequency   8Hz
Minimum pulse width   62.5ms
Contact resistance Event:   Active if resistance <35KΩ Inactive if resistance >200KΩ Status not defined if 35KΩ < resistance <200KΩ between input terminal and ‘C’ terminal

Serial Communications    
(Typical applications: Input of ASCII string inputs from Bar-code readers, credit card readers, Modbus etc.)

ASCII Printer      
ASCII Printer support    

Isolation†      
Terminals to ground:   100V RMS or dc (basic insulation)
† All isolation figures are:    
DC to 65Hz; BS EN61010 Installation category II; Pollution degree 2:

Transmission standard
EIA232 or EIA485
 
 

Last Updated Monday, July 18, 2005