- Home - Site Map - Links - Contact Us - About Us - Get In The Loop Search our site:  
5100V
Technical Specifications

TECHNICAL SPECIFICATION (Recorder)
Board types and hardware options
Six-channel universal input
Model 5180V: Eight boards (48 channels) max.
Model 5100V: Three boards (18 channels) max.
Three Change-over relay output board
Model 5180V: Nine boards (27 outputs) max.
Model 5100V: Four boards (12 outputs) max.
3.5 inch floppy disk, or PC Card (ATA flash or hard disk).
Environmental Performance
Temperature limits PC Card Option: Operation: 0 to 50°C; Storage: -25 to 70°C
Floopy disk drive option: Operation: 5 to 40°C; Storage: -20 to 50°C
Humidity limits Pc Card option: Operation: 8% to 85% RH; Storage: 8% to 90% (both non-condensing)
Floopy disk drive option: Operation 20% to 80% RH; Storage: 8% to 80% (both non-condensing)
Protection Bezel and display IP65
Shock BS EN61010
Vibration (10 to 150 Hz) 2g peak
Altitude <2000 metres.

Electromagnetic compatibility (EMC)
Emissions BS EN50081-2
Immunity BS EN50082-2

Electrical saftey
(BS EN61010) Installation cat. II; Pollution degree 2

Physical (Model 5100V)
Panel mounting DIN43700
Bezel size 144 x 144 mm.
Panel cutout dimensions 138 x 138 (both - 0 + 1 mm)
Depth behind bezel rear face 248 mm
Weight 3 kg
Panel mounting angle
Recorders with hard disk: Vertical panel only
Recorders with floppy disk: +15°
Other 5100V recorders: +45°

Physical (Model 5180V)
Panel mounting DIN43700
Bezel size 144 x 288 mm
Panel cutout dimensions 281 x 281 (both - 0 + 1 mm)
Depth behind bezel rear face 305 mm
Weight 7.5 kg
Panel mounting angle
Recorders with hard disk : Vertical panel only
Recorders with floppy disk: +15°
Other 5180V recorders: +45°

Operator interface
Type Colour TFT LCD witih cold cathode backlighting.
Fitted with resistive, analogue, toughened touch-panel
Size and resolution
Model 5100V: 1/4 VGA (320 x 240 pixels)
Model 5180V: SVGA (800 x 600 pixels)

Power requirements
Line voltage 47 to 63 Hz 85 to 265V
Power (Max) 60VA (Inrush current 36A)
Fuse type None

Ethernet communications
Electrical standard 10Mbs Ethernet. 10BaseT.
Transport protocol TCP/IP. Provision for File Transfer Protocol (FTP)

Technical Specification (Input board)
General
Input types dc Volts, dc millivolts,
dc milliamps (with shunt),
Thermocouple,
2/3 - wire RTD
Contact closure (not chan. 1) >60ms
Input type mix Freely configurable.
Maximum number of inputs 6 per board
Termination Edge connector / terminal block
Noise rejection (48 to 62 Hz) Common mode: >140dB (channel to channel and channel to ground
Series mode: >60dB.
Maximum common mode voltage 250 Volts continuous
Maximum series mode voltage 45 mV at lowest range;
12 Volts peak at highest range.
Insolation (dc to 65 Hz; BS EN61010) Installation cat II; Polution degree 2
Channel to channel: 300V RMS or dc (double insulation)
Channel to common electronics: 300V RMS or dc (double insulation)
Channel to ground: 300V RMS or dc (basic insulation)
Dielectric strength (BS EN61010) (1 minute type tests)
Channel to channel 2300 Vac
Channel to ground 1350 Vac
Insulation resistance >10MW at 500 V dc
Input impedance 38mV, 150 mV, 1 V rnages: >10MW;
10V range: 68.8 kW
Over voltage protection 50 Volts peak (150V with attenuator)
Open cirucit detection +57 nA max.
Recognition time 500 msec
Minimum break resistance 10 MW

Update/archive rates
Input/Relay-output sample rate 8 Hz
Display update 1 Hz
Archive sample-value Latest value at archive time
Trend/Display value Latest value at display update time

DC Input ranges
Shunt Externally mounted resistor modules
Additional error due to shunt 0.1% of input
Additional error due to attenuator 0.2% of input
Performance See table 1
Low
Range
High
Range
Resolution Maximum Error
(Instrument at 20°C)
Worst case temperature
performance
-8 mV 38mV 1.4mV 0.085% input + 0.073% range 80ppm of input der deg C
-30mV 150mV 5.5mV 0.084% input + 0.053% range 80ppm of input per deg C
-0.2 Volt 1 Volt 37mV 0.084% input + 0.037% range 80ppm of input per deg C
-2 Volts 10 Volts 370mV 0.275% input + 0.040% range 272ppm of input per deg C
Table 1 DC performance

Thermocouple data
Temperature scale ITS 90
Bias current 0.05 nA
Cold junction types Off, internal, external
CJ error 1°C max with inst. at 25°C
CJ rejection ratio 50:1 minimum
Remote CJ Via any user-defined channel
Upscale / downscale drive High, low or none selectable for each
thermocouple channell
Additional error: 0.01 °C(type.) if high or low selected
Types and ranges See table 2
T/C Type Overall range (°C) Standard Maximum linearisation
error
B 0 to + 1820 IEC584.1 0 to 400°C: 1.7°C
400 to 1820°C: 0.03°C
C 0 to + 2300 Hoskins 0.12°C
D 0 to 2495 Hoskins 0.08°C
E 0 to 2495 IEC584.1 0.03°C
G2 0 to 2315 Hoskins 0.07°C
J -210 to + 1200 IEC584.1 0.02°C
K -270 to + 1372 IEC584.1 0.04°C
L -200 to + 900 DIN43700:1985
(To IPTS68)
0.20°C
N -270 to + 1372 IEC584.1 0.04°C
R -50 to + 1768 IEC584.1 0.04°C
S -50 to + 1768 IEC584.1 0.04°C
T -270 to + 400 IEC584.1 0.02°C
U -200 to + 600 DIN43710:1985 0.04°C
NiMoNiCo -50 to + 1410 ASTM E1751-95 0.06°C
Platinel 0 to + 1370 Engelhard 0.02°C
Table 2 Thermocouple types and ranges

Resistance inputs
Ranges (including lead resistance) 0to 150W, 0 to 600W, 0 to 6kW
Influence of lead resistance Error=negligible;
Mismatch=1W/W
Temperature Scale ITS90
Accuracy and resolution See table 3
Low
Range
High
Range
Resolution Maximum error
(Instrument at 20°C)
Worst case temperature
performance
0W 150W 5mW 0.045% input + 0.110% range 35ppm of input per deg C
0W 600W 22mW 0.045% input + 0.065% range 35ppm of input per deg C
0W 6kW 148mW 0.049% input + 0.035% range 35ppm of input per deg C
Table 3 Resistance ranges - accuracy and resolution

RTD Type Overall range (°C) Standard Max linearisation error
Cu10 -20 to + 400 General Electric Co. 0.02°C
JPT100 -220 to + 630 JIS C1604:1989 0.01°C
Ni100 -60 to + 250 DIN43760:1987 0.01°C
Ni120 -50 to + 170 DIN43760:1987 0.01°C
Pt100 -200 to + 850 IEC 751 0.01°C
Pt100A -200 to + 600 Eurotherm Recorders SA 0.09°C
Pt1000 -200 to + 850 IEC 751 0.01°C
Table 4 RTD types and ranges

Relay outputs
Maximum switching power* 500VA or 60W
Maximum breaking current* 2 Amps within above power ratings
Maximum contact voltage* 250V within above power ratings
Isolation Contact to contact: 300V RMS or dc (double insulation)
Contact to ground: 300V RMS or dc (basic insulation)
Estimated life* 30,000,000 operations
Last updated December 21, 2003